Section IV. PREVENTIVE MAINTENANCE
3-8. Cleaning.
dipped in ethyl alcohol (item 1, table 1-4). A dry,
lint-free cloth should be used to clean contacts.
The removal of grease, dirt, dust and grime on
the test set is necessary for the proper operation
3-9. Lubrication.
of the test set. All items, except electrical con-
No lubrication of the test set is required at the
tacts, should be wiped with a clean, lint-free cloth
organizational level of maintenance.
Section V. OPERATIONAL CHECKOUT
3-10. Performance Checks.
made to the Corrective Action column to correct
the malfunction. The steps given in table 3-2 must
Performance checks are given in table 3-2. Per-
be performed in the sequence indicated, in order
formance checks are used to determine if the test
to ensure proper system build-up and provide an
set is operating within its operating limits. If the
immediate means of detecting a fault. Before
test set does not meet the requirements estab-
starting the performance checks given in table 3-
Iished in the Normal Indication column of table 3-
2, the Power ON/OFF switch (2, figure 2-1) must
be set to OFF.
Table 3-2. Performance Checks
Normal
Abnormal
Corrective
Action
Indication
Indication
Action
Step
NOTE
1
Power ON indi-
Power ON indica-
Replace Power
Set Power
tor (3) does not
ON/OFF
cater (3)
ON indicator
switch (2)
lights.
light.
(3), 1/2 AMP
to ON.
fuse (1), or
power cable if
defective. If
normal indica-
tion cannot be
obtained, refer
test set to
higher level of
maintenance.
2
Set meghom-
See paragraph
meter
indicator (14)
indicator (14)
2-8f. If
RANGE
reads O meg-
does not read
proper read-
SELECTOR
ohms (ZERO
zero meg-
ing cannot be
switch (17)
CALIBRATE
ohms.
obtained,
to ZERO
mark).
refer test set
CAL.
to higher
level of
maintenance.
3-5
|