TM 55-4920-325-14&P
Table 3-2. Performance Checks-Continued.
Abnormal
Normal
Corrective
Indication
Indication
Action
Action
Step
See paragraph
3
Set megohm-
indicator (14)
2-8f. If
indicator (14)
meter
proper read-
does not read
reads one meg-
RANGE
ing cannot be
one megohm.
ohm (MIDSCALE
SELECTOR
obtained,
CALIBRATE
switch (17)
refer test set
mark).
to MID-
to higher
SCALE
level of
CAL.
maintenance.
See paragraph
4
Set FUNC-
2-6f. If
INDICATOR (6)
INDICATOR (6)
TION
proper read-
does not read
reads zero pF.
SELECTOR
ing cannot be
zero pF.
switch (25)
obtained,
to CAP
refer test set
IND-ZERO
to higher
CAL and
level of
Capaci-
maintenance.
tance indi-
cator
RANGE
SELECTOR
switch (7)
to X1.
See paragraph
5
Set FUNC-
INDICATOR (6)
2-6f. If
INDICATOR (6)
TION
proper read-
does not read
reads value
SELECTOR
value stamped
ing cannot be
stamped on
switch (25)
on STANDARD
STANDARD
obtained,
to CAP
refer test set
IND HIGH
CAPACITOR
CAPACITOR
to higher
MMF name-
MMF name-
CAL and
plate (24).
level of
plate (24).
Capaci-
maintenance.
tance
RANGE
SELECTOR
switch (7)
to X50.
3-6
|