TM 55-4920-411-13&P
a. Performance Check.
(1) Position controls as listed below:
(a) POWER SWITCH to OFF.
(b) FUNCTION SWITCH to 0%.
(c) VAR FREQ ADJ SWITCH to 0.
(d) TEMP SWITCH to NORM.
(2) Connect test set to autotransformer, using
power cable supplied with unit and connect autotrans-
NOTE
former to a 115-V ac power source and adjust for 115-V
output.
If test set will not adjust on both 400 and
800 settings, try adjusting R32 (fig 4-4 in-
(3) Set POWER SWITCH to ON and allow at least
stead of R40. Adjusting R32 changes am-
15 minutes for warmup
bient readings. Take a new ambient read-
(4) Connect differential voltmeter to SIGNAL
ing before returning to para 4-10a.
OUTPUT (J2) pin E(+) and pin D(-), using two leads (B2).
b. Troubleshooting and Repair (Temperature Cir-
(5) Place probe end of thermometer adjacent to ter-
minals of J2. Wait 10 minutes for thermometer to sta-
cuit).
bilize.
(1) If output voltage is several volts high and unad-
justable, check TEMP OUTPUT potentiometer RI for
NOTE
continuity and broken or loose wires. R1 is a 2K lo-turn
Constantly check thermometer temper-
pot. Replace R1 (55, fig 4-12), if defective.
ature values to insure proper millivolt
(2) If there is no output voltage or output voltage
values.
cannot be adjusted, see Troubleshooting Table 4-3.
Check connector J2 for continuity between pins J and H
(6) Set TEMP OUTPUT control for 400C plus
and J and G with switch S3 in NORM position. Fig 4-5
value in degrees C as indicated on thermometer. If dif-
shows pin arrangement of J2.
ferential voltmeter indication is not between 16.14 and
16.65 mV dc, adjust R40 for 16.400 mV dc indication on
RESULT : Continuity shall be present.
differential voltmeter. If indication cannot be met, see
(3) If continuity is absent, open simulator and
check switch S3 for continuity (fig 4-5). Switch action is
(7) Set TEMP OUTPUT control for 800C plus
NORM: 2-1, 5-6; LOP: 1-2, 4-5; HI OIL TEMP: 2-3, 5-6.
value in degrees C as indicated on thermometer. If dif-
Replace S3 if defective. Also replace seal (54, fig 4-12), if
ferential voltmeter indication is not between 33.031 and
worn or damaged.
33.523 mV dc, perform (a) and (b) below. If necessary, see
para b and table 4-3.
(4) If S3 checks good, check wiring between S3 and
J2. Check connector J2 for continuity between pins J
(a) Adjust R42 for 33.277 mV dc indication on dif-
and H and between J and G when switch S3 is held in
ferential voltmeter.
LOP position
(b) Repeat (6) and (7), as necessary for best in-
RESULT: Continuity present from J to H and ab-
tolerance compromise.
sent from J to G.
(8) Repeat techniques of (7) above at TEMP OUT-
(5) If result is not true, open simulator and check
PUT control settings listed in table 1-8. Differential
switch S3 for continuity. Replace S3 if defective.
voltmeter indications wil1 be within limits specified
4-23
|