TM 55-4920-411-13&P
Table 4-3. Troubleshooting Table, AVIM (continued)
TROUBLE
PROBABLE CAUSE
REMEDY
Replace
Defective lamp, S4
All of lamps L1-L6 do not light when LAMP TEST
(Fig 4-1 and 4-2)
switch is pressed.
R10-R15 or CR1-CR4 on dis-
play board or associated com-
(Para 4-8)
ponents
LAMP TEST switch, S4
(Replace
None of lamps L1-L6 light when LAMP TEST
switch is pressed.
Repair
+ 12 vdc or + 5 vdc supply on
(Para 4-8)
signal generator board
Replace
R31 on signal generator board
(Fig FO-1)
(Para 4-8)
Monitor oscil -
Clock oscillator out of adjust-
Output frequencies out of tolerance.
lator frequency
ment
between test
points 1 ( + ) and
3 (-) (Figure 4-7)
and adjust R 10
for 1 microsec-
ond period
Replace
One of frequency signal gener-
No output frequency when FUNCTION SWITCH
ator components defective
is in any position other than 0% or output frequen-
(See Fig 4-7 and
cy out of tolerance and oscillator frequency cor-
FO-1).
rect.
Replace
T2, CR10, CR1l, A4, A5,
No temperature output
Replace
C27 on signal generator board
or associated components.
Replace
TEMP OUTPUT poten-
Temperature output several volts high and unad-
tiometer open (R1) or
justable.
associated components
4-5
|