TM 9-6625-2803-14&P
WATT/VAR METER
4.0
4.1
TYPE - The Watt/Var Meter is of the transducer type,
designed for operation with three-phase, four-wire circuits,
as well as two and three-wire single phase circuits.
4.2
SCALES - The Watt/Var Meter has a range of 0 to 1.2KW/KVAR,
corresponding to the 3-phase range on the 5.5A and 125V switch
positions.
Other ranges are obtained using internal or external
CT'S or PT's. The internal Potential Transformers are connected
to the "Select Power - Set - Range" switch. The CT inputs
are to the A, B and C load binding posts which are marked
with the current range. Use the full scale switch voltage and
the value of current determined from the binding posts to
determine the scale multiplier printed on the face of the
KW/KVAP Meter. Multiply the meter readings by this multiplier
(Note:
For single-phase, three-
to obtain correct KW/KVAR.
wire , the indication is half of the actual value).
4.3
KW/KVAR READINGS - EXAMPLE:
If the voltage setting is 250
on BOTH switches, and the load wires are in the 50A binding
posts, the scale multiplier is 20.
If the scale reading is .80,
then the power is .80 x 20 = 16KW, if the "Select Power" switch
is set on "3", or if the "Select Power" switch is set on lead
or lag KVAR, the indication is 16KVAR.
4.4
ACCURACY - The Watt/Var Meter indication is accurate
to 2% of full scale on any frequency from 45 to 425Hz.
4.5
CAUTION - Set the "Select Power - Set - Range" voltage
switch higher than the actual phase to phase voltage as read on
the Voltmeter, or damage can result. Observe maximum current
ratings in connecting to load terminals.