1.1.2.5
TEST
This mode provides the means for testing the majority of the internal logic circuitry of the
Flow Indicator. An internal clock referenced frequency, selected by the Multiplier switch, is
substituted for the input frequency. The Flow Indicator is in all other respects essentially in
the RATE mode. All readings in the TEST mode must be exact.
1.1.2.6
REMOTE
This position of the function switch transfers control of the mode, thumbswitch and multiplier
selection to external means through the REMOTE Connector. All selections are made by
either contact closures or TTL/DTL logic "0", i.e.+0.4 V max. current sinking, two TTL loads
max.
1.1.3
Controls (3.2.4)
Complete selection of control functions are provided so that the user has complete flexibility
in the application of the Flow Indicator per Specification MIS10391A.
1.1.3.1
Function Selector Switch
A six position rotary switch, on the front panel, is provided to allow the user to select any one
of the five operating modes or to transfer control to a remote location.
1.1.3.2
Preset Switches
Six decades of thumbswitches are provided on the front panel. A decimal point is indicated
at the extreme left, i, e. .XXXXXX.
1.1.3.3
Preset Multiplier
A four position rotary switch on the front panel, with X1, X10, X100 and X1000 capability, as
required by Section 1.1.2.
1.1.3.4
RESET Switch
A momentary pushbutton switch on the front panel. Resets all internal circuitry in
preparation for a new command. May not reset the display to zero if the Memory
1-6
|