TM 9-4910-736-14&P
The meter should indicate
(24)
Set the power supply for 12 V dc.
26.5 to 27.5 KV.
(25)
Disconnect the TEST LEADS.
The meter should indicate approximately
zero.
e.
Condenser Capacitance Test Circuits.
Proceed as follows:
(1)
Set FUNCTION switch to COND and if not already done, connect
BAT LEADS to 12 V dc supply.
(2)
Set CONDENSER switch to CAP.
Connect TEST LEADS together.
(3)
(4)
Adjust trimmer potentiometer R25 (figure FO-2) for 30 KV indication
on black scale.
Connect TEST LEADS across 0.5 uF test capacitor.
(5)
(6)
Adjust trimmer potentiometer R23 (figure FO-2) for 0.5 UF
indication on red scale.
Connect TEST LEADS across 0.1 uF test capacitor. The meter
(7)
should indicate 0.09 to 0.11 uF.
(8)
Set power supply for 16 V de.
The meter indication should be
0.09 to 0.11 uF.
Disconnect the TEST LEADS.
(9)
NOTE
Other test capacitors may also be used to calibrate at
other points along the scale.
f.
Ohmmeter Test Circuits. Proceed as follows:
(1) Set FUNCTION switch to OHM and if not already done, connect BAT
LEADS to 16 V dc supply.
(2)
Set OHMMETER RANGE switch to 0-15.
Connect the TEST LEADS together.
(3)
(4) Adjust ZERO ADJ OHM AND SERIES RES control for zero ohm
indication.
(5) Connect the TEST LEADS to 50 ohm test resistor. The meter should
indicate 4.7 to 5.3 ohms.
(6) Set OHMMETER RANGE switch to 0 - 100K.
3-20
|