Click here to make tpub.com your Home Page

Page Title: Condenser Test Circuits-continued
Back | Up | Next

Click here for thousands of PDF manuals

Google


Web
www.tpub.com

Home


   
Information Categories
.... Administration
Advancement
Aerographer
Automotive
Aviation
Construction
Diving
Draftsman
Engineering
Electronics
Food and Cooking
Logistics
Math
Medical
Music
Nuclear Fundamentals
Photography
Religion
   
   

 

TM 9-4910-736-14&P
Connect TEST LEADS and oscilloscope across 0.1 uF test
(5)
capacitor.
(6)
Adjust trimmer potentiometer R72 (figure FO-2) for 450 mV p-p
indication on oscilloscope.  Note the frequency of the signal.
Connect oscilloscope to junction of R99 and CR21 (E5) and
(7)
ground.
(8)
Adjust trimer potentiometer R67 for a clean sine wave twice the
frequency displayed in step (6) with an amplitude of approximately
150 mV p-p.
Disconnect the oscilloscope.
(9)
Connect the TEST LEADS together.
(10)
(11)
Verify that ZERO ADJ OHM AND SERIES RES control is at 12 o'clock
position and then adjust trimmer potentiometer R55 (figure FO-2)
for a full-scale deflection (30 KV black scale).
(12)
Connect TEST LEADS across 0.1 uF test capacitor.
(13)
Adjust trimmer potentiometer R68 for 30 KV indication on the
black scale.
(14)
Repeat steps (10) through (13) until meter indicates 30 KV
whether the TEST LEADS are shorted together or connected across
the 0.1 uF test capacitor.
(15)
Connect the TEST LEADS to 0.3 ohm test resistor.
(16)
Adjust trimmer potentiometer R83 (figure FO-2) until meter
indicates 27 KV.
(17)
Connect the TEST LEADS together.
(18)
Adjust trimmer potentiometer R55 for 30 KV indication.
(19)
Connect the TEST LEADS to 0.3 ohm test resistor.
(20)
Adjust trimmer potentiometer R83 (figure FO-2) for 27 KV indication.
(21)
Repeat steps (14) through (17) until meter indicates 30 KV with
TEST LEADS shorted and 27 KV when connected across 0.3 ohm test
resistor.
(22)
Connect the TEST LEADS across the 0.1 uF test capacitor wired in
series with 0.3 ohm test resistor.  Meter indication should be
26.5 to 27.5 KV.
Connect the TEST LEADS across 0.5 uF test capacitor wired in
(23)
series with 0.3 ohm test resistor. Meter should indicate 26.5 to
27.5 KV.
3-19

Privacy Statement - Press Release - Copyright Information. - Contact Us

Integrated Publishing, Inc. - A (SDVOSB) Service Disabled Veteran Owned Small Business