|
|
(a) Remove the reflectoscope cover.
(j) Turn the SYNC switch (on transigate)
to MAIN PULSE, and the GATE OFF switch
( b ) Plug the reflectoscope into a 115 volt
10 volt, 50 / 60 cycle single phase power
on.
source. Turn the POWER (SCALE ILLUM)
(k) Turn the START knob (on trans-
igate) to "2". A d j u s t t h e S T A R T V E R N I E R
knob clockwise and allow time for warm-up.
k n o b so that a downward jog is in line with the
(c) Turn the POWER (SCALE ILLUM)
7th from the left vertical scale line on the display
knob counterclockwise, but not to "OFF".
screen. Place the search fixture UPPER-
(d) Turn the SENSITIVITY control
LOWER switch to "UPPER".
knob to "l", an d the REJECT knob to "OFF".
(l) Turn the LENGTH knob (on trans-
( e ) The wave form on the display screen
igate) to "1". A d j u s t t h e L E N G T H V E R N I E R
should be similar to view G, fig. 6-53.
Note. Should the wave form not appear,
knob to end the gate at the right hand edge of the
o r is markedly different than that shown in view
f l a w signal (approximately 9-1 / 2 vertical scale
lines from the left of display screen).
G, figure 6-53, recalibrate the tester in ac-
cordance with the instructions in paragraph d,
( m ) Adjust the VERT screw to place the
lower gate line over the zero horizontal scale line
above.
on display screen.
(f) Pour water into the master checking
( n ) Adjust the FOCUS and ASTIG knobs
g a g e to the "FILL" level (five inches from top),
for the sharpest wave form.
refer to paragraph d (3) (a). Place the marked
(0) Adjust the SENSITIVITY control
defect portion of the gage facing the operator.
k n o b (Pulser / receiver unit) until the echo pulse
Note. Wipe all air bubbles from the
measures 2-1 / 2 inches high on the display
cylinder wall. Bubbles may cause a false alarm
screen (view D, fig. 6-53).
signal.
Turn
the
REJECT knob
(p)
( g ) Remove the protector from the search
(pulser / receiver unit) clockwise to the reduce
fixture probe. Place the search fixture in the
t h e echo pulse height to 1 inch (view E, fig. 6-
master checking gage with the fixture flange
53).
resting on the cylinder mounting flange. Move
t h e probe to the top of its travel and place the
(q) Turn the ALARM LEVEL knob (on
transigate) to + AUTO RESET. Rotate the
fixture UPPER-LOWER switch to "UPPER".
ALARM LEVEL vernier knob clockwise, then
(h) Connect the cable to the search fixture
c o u n t e r c l o c k w i s e until the light just appears on
and the reflectoscope at connector "R". The
RESET button. Place the AUDIO switch to
wave form on the display screen will be similar to
"ON" and the alarm will sound (buzz).
view H, fig. 6-53.
(r)
Turn
the R E J E C T k n o b ( o n
( i ) Rotate the search fixture handle clock-
wise to move the probe downward. Rotate the
pulser / receiver unit) counterclockwise to
search fixture handle until the echo pulse, or
" O F F " . The echo pulse height will then return to
2 - 1 / 2 inches (view D, fig. 6-53).
spike (view D, fig. 6-53) is shown on the display
screen when the fixture pointer alines with the
(S) Continue clockwise rotation of the
s e a r c h until handle until a heavy echo reflection
defect on the master checking gage.
enters the gate (view F, fig. 6-53). The heavy
(j) Rotate the search fixture handle to
m o v e probe slightly upward and downward until
r e f l e c t i o n indicates that the search fixture probe
a maximum pulse is seen on the display screen.
h a s reached the maximum downward travel. The
c y l i n d e r barrel tester is now adjusted to establish
(k) Adjust the SENSITIVITY control
the wave form.
knob to probe an echo pulse height of 2-1 / 2
inches (view I, fig. 6-53).
(t) Turn the POWER KNOB to "OFF".
Disconnect cable and remove the search fixture
(l) Turn the REJECT knob clockwise to
from the master checking gage. Install probe
reduce the echo pulse height to 1 inch.
protector and reflectoscope cover. Pour water
(m) Turn the ALARM LEVEL knob (on
from checking gage.
transigate) to + AUTO RESET. Rotate the
e. Establish the Wave From on the Reflec-.
ALARM LEVEL vernier knob until the light
toscope Display Screen for Daily Operation.
just appears on RESET button. The alarm will
(1) Preliminary calitiration.
s o u n d (buzz) when the AUDIO switch is "ON".
6-64
|
Privacy Statement - Press Release - Copyright Information. - Contact Us |