Click here to make tpub.com your Home Page

Page Title: PROCEDURE FOR ALTIMETER AND R ATE OF CLIMB TEST, NEGATIVE ALTITUDE.
Back | Up | Next

Click here for thousands of PDF manuals

Google


Web
www.tpub.com

Home


   
Information Categories
.... Administration
Advancement
Aerographer
Automotive
Aviation
Construction
Diving
Draftsman
Engineering
Electronics
Food and Cooking
Logistics
Math
Medical
Music
Nuclear Fundamentals
Photography
Religion
   
   

 

TM 55-4920-416-13
NAVAER 17-15C-539
Section IV
5,280 feet above mean sea level. In a circumstance of this
The airspeed indicator pointer will rotate clockwise indicating
sort, the altimeter needle will read 5,280 feet if the barometric
an increase in simulated airspeed.
adjustment is properly set.  Connect the test set to the
g.  Open the PRESSURE increase control valve 3 until
instruments to be tested as indicated in paragraphs 4-34 and
the airspeed indicator pointer stabilizes at the required check
point.
h.  Stabilize the airspeed indicator at the required check
4-44. PROCEDURE FOR ALTIMETER AND R ATE  OF
points by opening the PRESSURE decrease control valve 4.
CLIMB TEST, NEGATIVE ALTITUDE.
i.  After completing all check points listed on the
a.  Set barometric indicator on Altimeter to 29.92.
airspeed Indicator calibration card, throw power switch to OFF.
b.  Set barometric indicator on Altimeter being tested or
j.  Slowly o pen PRESSURE control valves 3 and 4 until
calibrated to 29.92.
all instruments return to normal.
c.  Set  PRESSURE  selector  valve  at  NEGATIVE
k.  Set PRESSURE selector valve at OFF.
ALTITUDE.
d.  Close PRESSURE control valves 3 and 4.
4-49. CALIBRATE AIRSPEED INDICATOR, 50-650 KNOTS.
e.  Set VACUUM selector at OFF.
Adjust PRESSURE control valves 3 and 4 in accordance with
f.  Open VACUUM control valves 1 and 2.
paragraph 4-48. Set the master airspeed indicator at each
g.  Throw power switch to ON.
point shown on the calibration card. Enter the readings of the
h.  Slowly open the PRESSURE increase control valve 3.
instrument under test on the calibration test card provided.
The altimeter pointer will rotate counterclockwise indicating a
decrease in simulated altitude. At the same time, the rate-of-
4-50. CASE LEAK TEST, AIRSPEED INDICATOR.
climb indicator will move into the DIVE region.
a.  Adjust pressure control valves in accordance with
paragraph 4-48 and stabilize the airspeed indicator at 500
knots.
CAUTION
b.  Close needle valve on hose assembly to instruments
Use extreme care when conducting this
being tested.
test. If the altimeter goes below -1,000
c.  Observe the airspeed indicator being tested; the
feet the altimeter will be damaged.  Do
pointer should not move more than 4 knots in 60 seconds. A
not exceed  6,000 fpm during this test or
reading in excess of this indicates a possible case leak and will
the  rate-of-climb  indicator  will  be
necessitate removal of the instrument.
damaged.
d.  Slowly open needle valve on hose assembly to
i.  Stabilize the altimeter at the required check points by
instruments being tested.
slowly opening the PRESSURE decrease control valve 4.
e.  Return  master  instruments  to  normal  position
j.  After comple ting all check points listed on the
according to instructions in paragraph 4-48.
altimeter calibration card, throw power switch to OFF.
k.  Slowly open PRESSURE control valves 3 and 4 until
4-51.  Remove test hose connections in accordance with
all instruments return to normal.
l.  Set PRESSURE selector valve at OFF.
4-52.  MAXIMUM ALLOWABLE AIRSPEED TEST.  To test
4-45.
CALIBRATE ALTIMETER, NEGATIVE ALTITUDE.
and calibrate the Maximum Allowable Airspeed point on
Adjust PRESSURE control valves in accordance with
Mach-Airspeed indicators, make use if the master panel
paragraph 4-44. Set master altimeter at each point shown on
ALTIMETER and AIRSPEED INDICATOR on the VPT-7A Test
calibration card down to and including -1,000 feet. Enter the
Set, and proceed as follows.
readings of the instrument under test on the calibration test
card provided.
NOTE
The MACH-AIRSPEED INDICATOR with
4-46.  Remove test hose connections in accordance with
MAXIMUM
ALLOWABLE
AIRSPEED
POINTER, to be tested and calibrated,
usually has two adjustments, one at the
4-47.
AIRSPEED INDICATOR TEST, 50-650 KNOTS.
front lower left hand corner and one at
Connect the test set to the instrument to be tested as indicated
the rear.  The front adjustment sets a
limiting stop for the maximum allowable
airspeed pointer, the rear adjustment sets
4-48. PROCEDURE FOR AIRSPEED INDICATOR TEST,
the critical Mach number for the aircraft
50-650 KNOTS.
on the rim of the main outer dial. Before
a.  Set PRESSURE selector valve at airspeed 0-650
proceeding with tests, make sure that
knots.
these  two  adjustments  have  been
b.  Close PRESSURE control valves 3 and 4.
properly set in accordance with data
c.  Set VACUUM selector valve at OFF.
supplied  for  the  aircraft  using  the
d.  Open VACUUM control valves 1 and 2.
instrument.
e.  Throw power switch to ON.
f.  Slowly open the PRESSURE increase control valve 3.
a.  Connect both the ALTIMETER-RATE-OF-CLIMB and
AIRSPEED quick connect nipples of the Test Set to the Static
and to the pressure lines respectively of the instrument
17

Privacy Statement - Press Release - Copyright Information. - Contact Us

Integrated Publishing, Inc. - A (SDVOSB) Service Disabled Veteran Owned Small Business