TM 55-4920-325-14&P
Table 2-1. Control Panel, Operating Controls, Indicators, and Connectors (continued).
Name of Control,
Function
Indicator, or
Index
Connector
Number
Megohmmeter ZERO ADJ
Permits calibration of
15
control (R107)-continued
MEGOHMMETER indicator
at zero setting when Meg-
ohmmeter RANGE SELECTOR
is set to ZERO CAL.
16
Megohmmeter MIDSCALE
Operates in conjunction with
ADJ control (R106)
Megohmmeter RANGE
SELECTOR switch. Per-
mits calibration of the
midscale setting when
Megohmmeter RANGE SELEC-
TOR is set to MIDSCALE CAL.
17
Megohmmeter RANGE
Operates in conjunction with
MEGOHMMETER indicator
SELECTOR switch
(S102)
to extend range of
resistance which can be
measured by MEGOHM-
METER indicator. When
set to MIDSCALE CAL or
ZERO CAL, permits
calibration of MEGOHM-
METER indicator at mid-
scale or zero points.
Provides for connection to
18
EXT RES terminals (E102
and E103)
external resistance to be
measured. Functions in
conjunction with
CAP-RES CHECK switch.
Provides means for connect-
TEST IND COAX connec-
19
ing test set to aircraft
tor (J401)
indicator coaxial connector.
Provides means for connect-
TEST IND UNSH connec-
20
ing test set to aircraft
tor (J402)
indicator unshielded con-
nector.
Provides means for connect-
TEST IND COMP con-
21
ing test set to aircraft
nector (J403)
indicator compensator
connector.
2-6
|