|
|
TM 9-6685-202-14
MI (C) 6685-202-14-3-8
Figure 3-8. Altimeter calibration-operating Setup.
with the vernier adjusting knob. Align the altitude index
valve completely and open the relief valve slightly to
on the vernier slide with the desired altitude mark on the
allow the pressure to rise to the proper point.
scale. The altitude is marked at the following intervals:
WARNING
-1,000 to 10,000 feet, marked each 100 feet,
The operator should not break the
10,000 to 0,000 feet, marked each 500 feet,
seal between the valve subassembly
50,000 to .0,000 feet, marked etch 1,000 feet.
and the barometer tube while
If unmarked altitudes are desired, position the vernier
vacuum Is applied.
slide to the pressure equivalent of the desired altitude.
(5) After the pressure has been adjusted to
(4) Open the vacuum valve slightly. The
align the mercury meniscus with the sighting ring, the
mercury will fall and the altimeters will indicate an
pressure in the test chamber will be the same as that for
increase in altitude. As the, pressure nears the desired
the altitude which the vernier slide was set. Read and
point, the meniscus will be inside the sighting ring but
record the altimeter readings.
will be visible through the open side of the ring.
b. Type A-1 Modified. The modified barometer
Continue to decrease the pressure until the highest point
does not have an altitude scale for direct comparison of
of the meniscus appears to barely touch the lower edge
altimeters. However, the modified barometer can be
of the sighting ring as in figure 3-5. If the pressure has
used for
been reduced too much as indicated by light being
visible above the entire meniscus, close the vacuum
3-15
|
Privacy Statement - Press Release - Copyright Information. - Contact Us |