Section VI.
FINAL INSPECTION PROCEDURES
a. Assemble the adapter tube assembly (1),
5-40.
GENERAL.
objective tube assembly (2), borescope
extension tubes (3), and illuminating head
The following inspection shall constitute the criteria for
assembly (4).
resolution. If the direct support instrument repair shop is
unable to repair the M3 Borescope to the extent that it
b. Position a steel scale (5) approximately 9-10
will meet the standard listed below, it must be forwarded
inches below and within the field of view of the
to depot for repair.
illuminating head assembly.
5-41.
RESOLUTION.
c. Turn the electric light control knob (6) to
maximum light.
NOTE
Borescope extension tubes I, II, and III must be
d. Focus the eyepiece assembly (7) in order to
attached to the objective tube assembly during
resolve 0.032 of the steel scale. See appendix G
resolution.
for converting decimal measurements into
fractions.
5-143/(5-144 blank)
|