3.3.4
(Cont'd)
C. Other Front Panel Controls (Cont'd)
INPUT MULTIPLIER - See Section 3.2.2 above.
MEM. SW - See Section 3.3.1, C.
COUNT ON/OFF Switch - Not used in this mode.
D. Rear Panel Connections
A (BNC) - Not used in this mode.
FLOWMETER Connector - Not used in this mode.
CONTROL Connector - See Section 3.3.1, D.
EXT. RESET SW or PULSE Pins - Provided for external reset as required. The SW terminal is
for mechanical or solid state closure to COMMON with a current requirement of 3 MA maximum
and a voltage drop of +0.4 V maximum. The PULSE terminal is for reset by a negative pulse.
See Section I for specification requirements.
GATE-SW Pin - Not used in this mode.
EXT. START Pin - Not used in this mode.
EXT. STOP Pin - Not used in this mode.
E. Rear Panel Controls
A/FLOW Switch - Not used in this mode.
EXT. T. B./INT. T. B. Switch - See Section 3.3.1,E.
3.3.5
PRESET COUNT
The PRESET COUNT mode of operation is actually a variation of a totalize mode. The input signal is
not fed directly to the totalizer and displayed as in a totalize mode, but is first prescaled (divided) by
the combined capacities of the thumbswitch preset system and the three-position multiplier toggle
switch. This prescaled input is then totalized and displayed by the counter. Control of the counting
process in this mode is identical to that for the TIME-EXT. (Interval) mode.
3-9