3.3.2
(Cont'd)
D. Other Front Panel Controls
POWER Switch - See Section 3.3.1,C.
RESET Pushbutton - See Section 3.3.1, C.
DISPLAY TIME Control - See Section 3.3.1,C.
INPUT MULTIPLIER - See Section 3.2.2 above.
MEM. SW - Engages (ON) or disengages (OFF), the memory logic. In ON position, the display is
updated automatically at the end of each count cycle ONLY, when in the particular operating mode
reset also occurs automatically. Switch must be in OFF when gating is by START-STOP SW or
STARTSTOP DC control signals as described above. In OFF, the display continually follows the
counting process.
COUNT ON/OFF Switch - Not used in this mode.
E. Rear Panel Connections
A (BNC) - The input signal for some of the TIME measurement modes is applied to this connector.
FLOWMETER Conn. - Not normally used in this mode.
REF.1 MHz BNC - Test point for calibration of internal oscillator.
CONTROL Connector - See Section 3.3.1,D.
EXT. RESET SW or PULSE Pins - See Section 3.3.1, D.
F. Rear Panel Controls
A/FLOW Switch - See Section 3.3.1,E. Not normally used in this mode.
EXT. T. B./INT. T. B. Switch - See Section 3.3.1,E.
3.3.3
TIME-EXT. (Interval) Measurements
This mode allows measurement of phenomena such as DC level changes, switch closures and pulse to
pulse spacings which would more easily be applied to the rear CONTROL connector. NOTE: All of
A. Front Panel Controls
POWER SW. - See Section 3.3.1, C.
RESET Pushbutton - See Section 3.3.1,C.