1.1.9.2
REMOTE PROGRAM
A rear mounted 50 pin connector is provided which allows complete programmability of
mode selection, and preset thumbswitch and multiplier number. Programming is by means
of single line connection in the case of mode and multiplier, and by 1,2,4,8 BCD connections
in the case of the thumbswitch number.
1.1.9.3
External Time Base
A rear mounted BNC connector is provided to allow the user to introduce an external
frequency within the range of 100 Hz to 1 MHz as the basic time base clock frequency. The
sensitivity of the input is better than 1 Volt pk-pk with an input impedance of greater than 1
Kilohm. Maximum input without damage is 10 Volts RMS.
1.1.9.4
CONTROL Connector
The following control functions and connections are provided through the CONTROL
connector:
1.1.9.4.1 CHASSIS Ground
Same as earth ground connected to 3rd wire of input power connector.
1.1.9.4.2 COMMON
Internal logic ground.
1.1.9.4.3 EXT. RESET PULSE
Requires a 4 V negative pulse, 0.5 minimum duration or 1 minimum rise
s
s
time.
1.1.9.4.4 EXT. RESET SW
Contact closure to COMMON resets the internal logic of the Flow Indicator in an
identical manner to the front panel RESET pushbutton.