Click here to make tpub.com your Home Page

Page Title: CALIBRATE RATE-OF-CLIMB INDICATOR (CLIMB-DIVE)
Back | Up | Next

Click here for thousands of PDF manuals

Google


Web
www.tpub.com

Home


   
Information Categories
.... Administration
Advancement
Aerographer
Automotive
Aviation
Construction
Diving
Draftsman
Engineering
Electronics
Food and Cooking
Logistics
Math
Medical
Music
Nuclear Fundamentals
Photography
Religion
   
   

 

TM 55-4920-416-13
Section IV
NAVAER 17-15C-539
j.  Stabilize the altimeter at the required check points by
NOTE
adjusting the VACUUM decrease control valve 1.
k.  Stabilize the rate-of-climb indicator pointer at the
When tests have been completed, restore
required check points in the DIVE region by adjusting the
the original connections in the aircraft
VACUUM decrease control valve 1.
making sure that they are adequately
l.  After completing all check points on the altimeter and
tightened so that they will not leak or
rate-of-climb indicator, throw power switch to OFF.
vibrate loose in flight.
m. Slowly open VACUUM control valves 1 and 2 until all
4-35.  For tests of the complete aircraft pitot system, the
instruments return to normal.
VPT-7A is connected to the pitot tube itself by means of the
n.  Turn VACUUM selector valve to OFF.
test hoses and special clamps and sleeves provide.  (See
o.  During this test, disregard any movement on the
figure 4 -4.) Connection procedures are as follows:
airspeed indicator.
a.  Select the appropriate clamp, round or oblong, and fit
it carefully over the pitot tube so that the hole leading to the
4-39.  CALIBRATE RATE-OF-CLIMB INDICATOR (CLIMB-
clamp fitting is in line with the static hole on the pitot tube, and
DIVE). Adjust VACUUM control valves 1 and 2 in accordance
tighten clamp securely.
with paragraph 4-38. Set the master rate-of-climb indicator at
b.  Attach suitable length of test hose from the clamp to
each point shown on the rate-of-climb indicator calibration
the quick-connect nipple marked ALTIMETER AND RATE OF
card. Enter the readings of the instrument under test on the
CLIMB.
calibration test card provided. If necessary, adjust the rate-of-
c.  Select appropriate sleeve, round or oblong, fit
climb indicator to exactly zero before beginning this test.
adjustable hose camp on sleeve, and push firmly over the end
of pitot tube. Tighten clamp.
4-40.  CALIBRATE ALTIMETER, 0 - 50,000 FT AND 0 -
d.  Connect appropriate length of test hose to sleeve and
80,000 FT. Adjust VACUUM control valves In accordance with
to quick-connect nipple marked AIRSPEED. If single hose is
paragraph 4-38. Set the master altimeter at each point shown
too short, couple two together.
on the calibration card up to and including 50,000 feet or
e.  At conclusion of tests do not fail to remove sleeve and
80,000 feet depending on altimeter. Enter the readings of the
clamp.
instrument under test on the calibration test card provided.
4-36. OPERATIONAL  TESTS
AND
CALIBRATION
OF
4-41. CASE
LEAK
TEST
(RATE-OF-CLIMB
AND
INSTRUMENTS.
ALTIMETER).
a.  If necessary, adjust the pointer to read exactly zero
4-37. ALTIMETER AND RATE OF CLIMB TEST, 0 - 50,000
on the rate-of-climb indicator being tested.
FT AND 0 - 80,000 FT. Connect the test set to the Instruments
b.  Adjust VACUUM control valves in accordance with
to be tested as indicated in paragraphs 4-34 and 4-35.
paragraph 4-38 and stabilize the altimeter at 18,000 feet.
c.  Close needle valve on hose assembly to instruments
4-38. PROCEDURE FOR ALTIMETER AND RATE-OF-
being tested.
CLIMB TEST, 0 - 50,000 FT AND 0 - 80,000 FT.
d.  Observe the altimeter being tested.  The pointer
a.  Set barometric indicator on Altimeter to 29.92.
should not move more than 100 feet in 10 seconds. A reading
b.  Set barometric indicator on instrum ent being tested or
in excess of this indicates a possible case leak and will
calibrated to 29.92.
necessitate removal of the instrument.
c.  Set PRESSURE selector valve at OFF.
e.  Observe the rate-of-climb indicator being tested. The
d.  Open PRESSURE control valves 3 and 4.
point should not indicate more than 60 fpm DIVE.  Any
e.  Set VACUUM selector at ALT. & R.O.C. 0 - 50,000 FT
readings in excess of this indicate a possible case leak and will
or 0 - 80,000 FT for 80,000 foot altimeter.
necessitate the removal of the instrument being tested.
f.  Close VACUUM control valves 1 and 2.
f.  Slowly open needle valve on hose assembly to
g.  Throw power switch ON.
instruments being tested.
h.  Slowly open the VACUUM increase control valve 2.
g.  Return instruments to normal position accordin g to
The altimeter pointer will, rotate clockwise indicating an
instructions in paragraph 4-38.
increase in simulated altitude. At the same time, the rate-of-
climb indicator will move into the climb region.
4-42.  Remove test hose connections in accordance with
i.  Open the VACUUM Increase control valve 2 until the
rate-of-climb indicator pointer stabilizes at the required
checkpoint.
4-43. ALTIMETER AND RATE OF CLIMB TEST NEGATIVE
ALTITUDE.  When calibrating altimeters, at a considerable
height above sea level, it is necessary to provide pressure to
CAUTION
simulate sea-level or below sea-level indications on the
altimeter.
For  instance,  NAS  Denver,  Colorado,  is
Do not exceed  6,000 fpm during this test
as damage will occur to the rate-of-climb
indicator.
16

Privacy Statement - Press Release - Copyright Information. - Contact Us

Integrated Publishing, Inc. - A (SDVOSB) Service Disabled Veteran Owned Small Business