TM 55-4920-416-13
Section I
NAVAER 17-15C-539
provided are used for the power sources specified, the motor is
1-8. PERFORMANCE.
automatically connected for proper operation.  This feature
prevents connection of the test set to the wrong power source.
1-9. Table 1-2 lists supply requirements, performance data
and other leading particulars.
1-14. VACUUM AND PRESSURE SYSTEM (Figure 1 -5).
TABLE 1-2
1-15.  The principal components of the vacuum-pressure
system are as follows:
PERFORMANCE DATA
a.  A compact dual-purpose motor-pump assembly (1
and 2, figure 1-6). The motor section of the assembly is a
Type.......................................................Vacuum-Pressure Test Set
12,000 rpm, 1/30 horsepower, continuous duty, series wound
Model.......................................................................................VPT-7A
28 V dc motor. The pump section is a direct drive, positive
Power Requirements ...............................................28 V dc at 2.2A
displacement, rotary type pump with multiple porting.
115 V ac 1 ∅ 50-500 cps
b.  An air-oil reservoir (3, figure 1-6), connected to the
115 V ac 3 ∅ 50-500 cps at 0.75A rms
pressure side of the pump, feeds filtered air under pressure to
Operating Temperatures ....................................... -40F to  150F
the pressure control valves (4, figure 1-7). The vacuum side of
(-40C to  65C)
the pump is connected to the vacuum control valves (3,
Max Pressure Produced ....................................................50 lb/in.2
Max Vacuum Produced.................................................-29.5 in. Hg
c.  Two banks of relief valves (1 and 2, figure 1-7); three
Net Weight............................................................................57-1/2 lb
for vacuum and three for pressure. Their function is to protect
Shipping Weight Minus Overseas Packing............................67 lb
the master instruments (8, 9, 10, 11, 12, figure 4-1) and aircraft
Oil Sump Capacity..............................................................8 fluid oz.
instruments under test from excessive pressure or vacuum.
Dimensions
d.  The vacuum and pressure selector valves (1 and 2,
Depth ..........................................................................14-1/32 in.
figure 4-1). These selector valves switch multiple pneumatic
Height...........................................................................14-3/4 in.
circuits. For each selector valve test position all necessary
Width ............................................................................16-1/2 in.
inlet and outlet, pressure and vacuum lines, control valves and
Storage Temperatures ......................................... -54C to  84C
instruments are properly connected.
(-65F to  183F)
e.  A bank of simple plug valves (1, 2, 3, 4, 5, figure 4-2).
Pump Speed ................................................................. 12,000 RPM
These plug valves are located under the calibration card cover.
Rate-of-Climb Indicator ..................................0-6,000 ft/min (max)
They provide means for checking and calibrating master panel
climb or descent
instruments in the tester and are not used during routine
Altimeter Range.................................................. -1,000 to 80,000 ft
testing.
Airspeed Indicator Range .......................................... 50-650 knots
Manifold Pressure Indicator Range..............................10 to 75 in.
1-16. THEORY OF OPERATION (Refer to figure 1-5).
of Hg Absolute
Fuel Pressure Gauge Range .........................................0-25 lb/in.2
1-17. PRESSURE SECTION. The pump (1) serves the dual
Life .......................................................................................2,000 Hrs
function of developing pressure as well as vacuum for the
operation of the test set. In the pressure section a mixture of
1-10. DESCRIPTION OF THE MAJOR COMPONENTS.
air and oil is pumped into the reservoir (2) where the oil and air
are separated. The separated oil drops to the bottom of the
1-11. GENERAL. The VPT-7A is composed of a number of
reservoir where some is fed to the pump for lubrication
major components. These are grouped into two systems for
purposes, the air being forced out at the top of the reservoir
convenience and ease of analysis. They are as follows:
under pressure. The pressurized air, after going through a
a.  Electrical system.
check-valve (3), whose function it is to prevent oil from entering
b.  Vacuum system and pressure system.
instrument lines during transportation, enters an oil sump (4)
Minor components, not discussed in paragraphs 1-13 to 1-14
and passes through a filter which Is an integral part of the
are subordinate to one of these systems and are discussed at
sump. Air from the sump flows through two pressure control
an appropriate place in the handbook.
valves (5) and (6). The pressure increase valve (C3) controls
the amount of air permitted in the system.  The pressure
1-12. ELECTRICAL SYSTEM (Figures 1-3 and 1-4).
decrease valve (6) opens the line to the outside ambient air
allowing system pressure to bleed off. Proper setting of these
1-13. The electrical system is designed for a wide range of
two valves enables the operator to maintain a pressure at any
power inputs. The tester will operate on 28 V dc or 115 V ac,
desired level within the limitations specified in Table 1-2. The
50 to 500 cycles, single phase or three phase line.  Major
available pressure level, as fixed by the control valves, is sent
components of the system are a step-down transformer (1,
to the pressure relief valve PR-3 (10), and pressure selector
figure 1-3), selenium rectifier (2, figure 1-3) and an electrolytic
valve (7). Setting the pressure selector valve at the desired
capacitor (3, figure 1-3). The electrical system is energized
test position then completes the circuit to the instrument under
through a multi-pin AN connector (4, figure 1-3) and power
test and simultaneously connects the instrument under test
switch (7, figure 4-1). The internal connections of the power
and the master Instrument to one of pressure relief valves
cables and test set are such that when the special cables
4